Characterization of BG28 and KG3 filter glass for Drive Diagnostic Attenuators
[摘要] BG28 and KG3 filter glasses were tested for use as attenuators in the NIF drive diagnostic (DrD) systems. Tests were performed in the Optical Sciences Laser facility with a 351 nm, 2-step, 3-nsec pulse at fluences ranging up to {approx} 1 J/cm{sup 2}. Single-shot measurements showed no solarization when the samples were allowed to relax for a week after exposure. KG3 filters exhibited no luminescence and no transient pulse distortion. BG28 filters luminesced appreciably and imposed a 'droop' (similar to 'square-pulse distortion') on the signals. The droop parameter is estimated at 0.50 {+-} 0.11 cm{sup 2}/J. Droop is explained in terms of known copper-doped-glass spectroscopy and kinetics (buildup of triplet-state populations, with excited-state absorption). Simulation of the distortion ({approx}1.6%) expected on a 1.8 MJ Haan pulse led to a minor redesign of the Drive Diagnostic with reduced fluence on the BG28 filters to reduce the droop distortion to 0.5%.
[发布日期] 2007-11-30 [发布机构] Lawrence Livermore National Laboratory
[效力级别] [学科分类]
[关键词] Luminescence;Absorption;Buildup;Kinetics;Simulation [时效性]