Inductively Coupled Plasma Mass Spectrometry Uranium Error Propagation
[摘要] The Hazards Control Department at Lawrence Livermore National Laboratory (LLNL) uses Inductively Coupled Plasma Mass Spectrometer (ICP/MS) technology to analyze uranium in urine. The ICP/MS used by the Hazards Control Department is a Perkin-Elmer Elan 6000 ICP/MS. The Department of Energy Laboratory Accreditation Program requires that the total error be assessed for bioassay measurements. A previous evaluation of the errors associated with the ICP/MS measurement of uranium demonstrated a {+-} 9.6% error in the range of 0.01 to 0.02 {micro}g/l. However, the propagation of total error for concentrations above and below this level have heretofore been undetermined. This document is an evaluation of the errors associated with the current LLNL ICP/MS method for a more expanded range of uranium concentrations.
[发布日期] 2001-07-01 [发布机构] Lawrence Livermore National Laboratory
[效力级别]  [学科分类] 
[关键词] Lawrence Livermore National Laboratory;Uranium;37 Inorganic, Organic, Physical And Analytical Chemistry;Mass Spectrometers;Bioassay [时效性] 