Compact Gamma-Ray Imager for In-Vivo Gene Imaging
[摘要] A compact, low-cost, gamma-ray imaging system is needed to study gene expression in small animals. State-of-the-art electronic imaging systems have insufficient resolution and animals must be sacrificed for detailed imaging that precludes time evolution studies. With improved electronics radioactive tracers attached to gene markers can be used to track the absorption and mobility of gene therapy medications in live animals. Other instrumentation being developed for medical applications does not have the response to match the radiation source for this work. The objective of this research was to develop thick film (Cd,Zn)Te detectors matched to the gamma ray energy of {sup 129}I. The detector would be a direct readout device using p-i-n diodes formed from the high Z material absorbing the radiation, with separate readout. Higher quality semiconducting material was expected from epitaxial growth on GaAs, a near lattice matched substrate. In practice, it was difficult to obtain material with high resistance and low leakage current. Spire Corporation achieved the goal of fabricating working detectors in (Cd,Zn)Te deposited on GaAs. The spectra of an alpha emitter ({sup 225}Am) was adequately resolved in thin film devices. Thick p-i-n diodes were fabricated but other processing problems prevented full demonstration of a gamma ray detector.
[发布日期] 2000-06-01 [发布机构] Spire Corporation, Bedford, MA (United States)
[效力级别]  [学科分类] 
[关键词] Spectra;Radiation Sources;Absorption;Leakage Current;Processing [时效性] 
   浏览次数:20      统一登录查看全文      激活码登录查看全文