Toward an optimized PET reconstruction protocol for spinal cord: superiority of OSEM with TOF and PSF
[摘要] BackgroundAlthough multiple PET reconstruction approaches have been successfully applied in brain and oncologic imaging, their performance in the context of spinal cord imaging remains unverified. This study aims to identify the optimal reconstruction method for spinal cord PET image analysis based on assessments of image quality and quantitative uptake.MethodsWe retrospectively analyzed 18F-fluorodeoxyglucose (FDG) PET/MR data from 22 patients with cervical spinal cord tumors. PET images were reconstructed using Ordered Subset Expectation Maximization (OSEM), OSEM + Point Spread Function (PSF), OSEM + Time-of-Flight (TOF), and OSEM + TOF + PSF algorithms. Image quality was assessed by noise, signal-to-noise ratio (SNR), and signal-to-background ratio (SBR), both overall and within equally sized small (n = 11) and large (n = 11) tumor subgroups. Tumor uptake metrics, including SUVmean, SUVmax, SUVRmean, and SUVRmax of the lesions, were quantified for each method. The differences among reconstruction methods were assessed using the Friedman test.ResultsThere were no significant differences in image noise among reconstruction methods (P > 0.05). Compared to OSEM reconstruction, the OSEM + PSF + TOF method significantly improved SNR (Z = 3.62, P < 0.01) and SBR (Z = 2.80, P = 0.03); the SBR of OSEM + PSF + TOF was also significantly higher than that of OSEM + PSF (Z = 3.04, P = 0.01). In large tumors, while image noise remained unaffected by the reconstruction method (P > 0.05), both SNR and SBR were significantly impacted (both P < 0.05). Specifically, OSEM + TOF + PSF achieved significantly higher SNR compared to OSEM (Z = 3.30, P < 0.01), and higher SBR compared to OSEM + PSF (Z = 3.14, P = 0.01). Quantitative lesion uptake metrics (SUVmax, SUVmean, SUVRmax, and SUVRmean) were significantly higher for OSEM + TOF + PSF compared to both OSEM and OSEM + PSF reconstructions (all P < 0.05). Additionally, SUVmean (Z = 2.92, P = 0.02) and SUVRmean (Z = 2.83, P = 0.03) values from OSEM + TOF + PSF reconstruction were significantly higher than those from OSEM + TOF.ConclusionThe OSEM + PSF + TOF reconstruction method demonstrated significant advantages in both image quality and quantitative assessment of FDG uptake, making it the recommended choice for spinal cord PET image reconstruction.
[发布日期] 2026-08-16 [发布机构] 
[效力级别]  [学科分类] 
[关键词] Hybrid PET/MR;Image reconstruction;Time-of-flight;Point-spread function;Spinal cord tumor [时效性] 
   浏览次数:1      统一登录查看全文      激活码登录查看全文