A large area high resolution imaging detector for fast atom diffraction
[摘要] We describe a high resolution imaging detector based on a single 80 mm micro-channel-plate (MCP) and a phosphor screen mounted on a UHV flange of only 100 mm inner diameter. It relies on standard components and we describe its performance with one or two MCPs. A resolution of 80 pm rms is observed on the beam profile. At low count rate, individual impact can be pinpointed with few pm accuracy but the resolution is probably limited by the MCP channel diameter. The detector has been used to record the diffraction of fast atoms at grazing incidence on crystal surfaces (GIFAD), a technique probing the electronic density of the topmost layer only. The detector was also used to record the scattering profile during azimuthal scan of the crystal to produce triangulation curves revealing the surface crystallographic directions of molecular layers. It should also be compatible with reflection high energy electron (RHEED) experiment when fragile surfaces require a low exposure to the electron beam. The discussions on the mode of operation specific to diffraction experiments apply also to commercial detectors.
[发布日期] 2018-07-15 [发布机构]
[效力级别] Proceedings Paper [学科分类]
[关键词] Fast atom diffraction;Reflexion high energy electron diffraction;Imaging detector [时效性]