MATERIALS IMPURITY ANALYSIS BY MEANS OF NUCLEAR-RESONANCE REACTIONS
[摘要] An automated beam line for measurement of trace impurities in materials is installed on the FN tandem in the Center for Accelerator Mass Spectrometry, LLNL. The major purpose of the beam-line system is to determine the diffusion of water into glasses by use of the F-19(p, alpha-gamma)O-16 reaction. This study is stimulated by the radiation-waste-isolation program; however, the system is applicable to studies of other impurities in many material matrices. The target wheels, each of which holds twenty-four samples, can be changed in less than an hour. The beam-line control system both indexes the target wheel and controls the tandem energy as it steps through the requested measurement protocol. The tandem control system, in response to a request, adjusts the terminal potential and other transport elements to maintain constant spot size and location.
[发布日期] 1991-05-01 [发布机构]
[效力级别] Proceedings Paper [学科分类]
[关键词] [时效性]