THE FEASIBILITY OF UTILIZING THE SYSTOLIC PRE-EJECTION INTERVAL AS A DETERMINANT OF PACING RATE
[摘要] Rate-modulated pacing modes adjust the stimulus rate by responding to sensed alterations in physiologic indexes of metabolic demand. This study was designed to determine whether right ventricular pre-ejection interval, measured in patients by a prototype pacemaker system capable of tracking intraventricular volume, changes predictably with exercise and, if so, whether it can be used in algorithm to vary heart rate appropriately. This system utilizes intraventricular electrical impedance mesurements of injected microampere currents to determine intracavitary volume changes. Five pacemaker-dependent patients underwent temporary insertion of a tripolar electrode connected to an external device that sensed cardiac signals, generated an impedance wave form and produced stimuli at rates dependent on pre-ejection interval. Pre-ejection interval did not change as a result of variations in pacing rate itself (347 .+-. 41 ms at 70 beats/min versus 321 .+-. 19 ms at 130 beats/min), but consistently decreased during graded exercise (by 23% from baseline). During rate-modulated pacing based on pre-ejection interval, heart rate significantly increased during exercise compared with ventricular demand pacing (by 46 .+-. 6 versus .+-. 6 beats/min, respectively), and increased appropriately during burst exercise. Thus, the pre-ejection interval appears to be a specific, reliable physiologic determinant of pacing rate during exertion, which may be applicable in implantable rate-modulated pacemakers.
[发布日期] 1989-12-01 [发布机构] 
[效力级别]  [学科分类] 
[关键词]  [时效性] 
   浏览次数:3      统一登录查看全文      激活码登录查看全文