DETERMINATION OF RIGHT VENTRICULAR PRESSURE IN THE PRESENCE OF A VENTRICULAR SEPTAL-DEFECT USING CONTINUOUS WAVE DOPPLER ULTRASOUND
[摘要] Continuous wave Doppler ultrasound was employed in 38 patients with ventricular septal defects, many with associated lesions, to measure the velocity (V) of the shunted blood. Using the modified Bernoulli equation (.DELTA.P = 4V2) the pressure difference (.DELTA.P) between the ventricles was determined. In 22 patients both right ventricular and either left ventricular or ascending aortic pressure were measured at the time shunt velocity was determined. In another 16 patients these measurements were not obtained simultaneously but in most they were done within 24 hours of each other. In the entire group, measured pressure differences between the ventricles (or aorta and right ventricle) ranged from 0 to 97 mm Hg (mean 52 .+-. 24). On the basis of velocity measurements the pressure difference ranged from 7 to 112 mm Hg (mean 51 .+-. 24). A close correlation was found between the two methods (r = 0.95, SEE = 7.8 mm Hg). This accuracy was not altered by associated lesions. These findings indicate that by the use of continuous wave Doppler interrogation right ventricular pressure can be accurately measured in the presence of a ventricular septal defect.