ARRHYTHMIAS INDUCED BY DEVICE ANTITACHYCARDIA THERAPY DUE TO DIAGNOSTIC NONSPECIFICITY
[摘要] New technology has produced automatic cardioverter-defibrillators capable of delivering antitachycardia pacing, as well as low and high energy shocks and backup bradycardia pacing. These expanded treatment options have led to a wider range of clinical applications for such devices, including the treatment of ventricular tachycardias with longer cycle lengths, which may overlap the cycle lengths of some supraventricular arrhythmias. The diagnostic capability of these devices, although improved, has not advanced sufficiently to ensure reliable discrimination between all supraventricular and ventricular arrhythmias. Two cases are presented in which device-mediated pacing therapy, triggered by supraventricular arrhythmias, induced ventricular tachycardia requiring additional therapeutic intervention. This report illustrates the therapeutic versatility and some of the potential pitfalls, of the recently developed devices and reviews the status of automatic arrhythmia identification technology.
[发布日期] 1991-11-01 [发布机构] 
[效力级别]  [学科分类] 
[关键词]  [时效性] 
   浏览次数:1      统一登录查看全文      激活码登录查看全文