ATRIAL SENSING PERFORMANCE OF THE SINGLE-LEAD VDD PACEMAKER DURING EXERCISE
[摘要] Objectives. The purpose of this study was to examine the atrial sensing performance of the single-lead VDD pacing system during exercise and concomitant changes in the amplitude of the atrial electrogram. Background. Studies of conventional dual-chamber pacing have demonstrated an overall reduction in the atrial signal amplitude and a variable incidence of atrial undersensing during vigorous exercise. Methods. The telemetered atrial electrogram and simultaneous surface electrocardiogram (ECG) were continuously recorded in 12 patients (mean age 70.8 years) with an implanted single-lead VDD pacing system during treadmill stress testing. The atrial signal amplitude was measured at peak exercise, and the patients were monitored for maintained atrial synchronized ventricular pacing during the entire exercise and recovery period. Results. The atrial electrographic voltage decreased an average of 19.5% (p < 0.05) during peak exercise, and the reduced P wave amplitude ranged from 6.9% to 59.4% of the preexercise values in 8 of 12 patients. Three patients showed a modest increase in atrial signal amplitude, and one patient had no change. The telemetered electrogram displayed persistent and intact atrial synchronous ventricular pacing throughout the study period. Conclusions. Despite relatively low atrial signal amplitudes at rest and further decreases during exercise, the single-lead VDD pacemaker maintains reliable atrial tracking and ventricular pacing during vigorous exercise.
[发布日期] 1993-12-01 [发布机构] 
[效力级别]  [学科分类] 
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文