Lead preconcentration onto C-18 minicolumn in continuous flow and its determination in biological and vegetable samples by flame atomic absorption spectrometry
[摘要] A procedure for the preconcentration and determination of lead in vegetable and biological samples was developed in the continuous mode coupled to a flame atomic absorption spectrometer. Lead is quantitatively preconcentrated in acetic buffer as its diethyldithiocarbamate chelate onto a C-18 minicolumn, placed in the loop of a proportional injector, eluted by a stream of methyl isobutyl ketone and introduced directly into the nebuliser. A detection limit of 3 mu g l(-1) is obtained using a time-based technique for 2 min preconcentration and an RSD of 3.8% was readily achieved for three measurements of 25 mu g Pb l(-1). The sample throughput is 24 h(-1). Using preconcentration times of 10 min an enrichment factor of 189 can be obtained. The continuous flow system was used for some reference sample analysis and the obtained results reveal that the methodology can be easily applied for vegetable and biological sample analysis.
[发布日期] 1996-06-01 [发布机构] 
[效力级别]  Proceedings Paper [学科分类] 
[关键词] biological samples;flame atomic absorption spectrometry;lead;preconcentration;vegetable samples [时效性] 