Spectrophotometric flow system using vanadomolybdophosphate detection chemistry and a liquid waveguide capillary cell for the determination of phosphate with improved sensitivity in surface and ground water samples
[摘要] A now injection procedure based on the vanadomolybdate method for the determination of dissolved reactive phosphorus in water samples is described. The system includes a liquid waveguide long pathlength cell connected to a charge Coupled device (CCD) spectrophotometer. Calibration was linear up to 500 mu gPL(-1), with a detection limit of 17 mu gPL(-1) and a quantification limit of 56 mu gPL(-1). An injection throughput of 60 determinations h(-1) and repeatability (R.S.D.) of 2.2% were achieved, Potential interference from silicate was effectively masked by addition of a tartaric acid stream. The accuracy of the proposed methodology was assessed through analysis of a certified reference material and recovery tests on water samples. The developed procedure allows the determination of phosphorus in water samples at trace levels with high sensitivity, reduced reagent consumption and low waste production. (C) 2008 Elsevier B.V. All rights reserved.
[发布日期] 2008-12-15 [发布机构]
[效力级别] Proceedings Paper [学科分类]
[关键词] Phosphate determination;Flow injection;Liquid waveguide capillary cell;Long pathlength spectrophotometry;Silicate interference [时效性]