Fluorimetric determination of fluoride in a flow assembly integrated on-line to an open/closed FIA system to remove interference by solid phase extraction
[摘要] A simple flow injection fluorimetric method for fluoride determination is proposed. The method is based on the enhanced fluorescence of quercitin-Zr(IV) complex when fluoride ion is present in the sample. An open/closed FIA manifold with a mini-column of Dowex 50W X8 resin was used to remove the most important interference (aluminum). The two FIA assemblies were integrated on-line to automate the pretreatment of the water sample and fluoride determination. The calibration graph was linear over the range 0.1-3.0 mug ml(-1) of fluoride with a correlation coefficient of 0.999 and LOD 0.06 mug ml(-1). The relative standard deviation was 2.5% and the sample throughput was 52 h(-1) without pretreatment and 10 h(-1) with pretreatment of the sample. The method was applied to the determination of fluoride in water samples. (C) 2002 Elsevier Science B.V. All rights reserved.