已收录 273124 条政策
 政策提纲
  • 暂无提纲
Comprehensive Cardiovascular magnetic resonance of myocardial mechanics in mice using three-dimensional cine DENSE
[摘要] BackgroundQuantitative noninvasive imaging of myocardial mechanics in mice enables studies of the roles of individual genes in cardiac function. We sought to develop comprehensive three-dimensional methods for imaging myocardial mechanics in mice.MethodsA 3D cine DENSE pulse sequence was implemented on a 7T small-bore scanner. The sequence used three-point phase cycling for artifact suppression and a stack-of-spirals k-space trajectory for efficient data acquisition. A semi-automatic 2D method was adapted for 3D image segmentation, and automated 3D methods to calculate strain, twist, and torsion were employed. A scan protocol that covered the majority of the left ventricle in a scan time of less than 25 minutes was developed, and seven healthy C57Bl/6 mice were studied.ResultsUsing these methods, multiphase normal and shear strains were measured, as were myocardial twist and torsion. Peak end-systolic values for the normal strains at the mid-ventricular level were 0.29 ± 0.17, -0.13 ± 0.03, and -0.18 ± 0.14 for Err , Ecc , and Ell , respectively. Peak end-systolic values for the shear strains were 0.00 ± 0.08, 0.04 ± 0.12, and 0.03 ± 0.07 for Erc , Erl , and Ecl , respectively. The peak end-systolic normalized torsion was 5.6 ± 0.9°.ConclusionsUsing a 3D cine DENSE sequence tailored for cardiac imaging in mice at 7 T, a comprehensive assessment of 3D myocardial mechanics can be achieved with a scan time of less than 25 minutes and an image analysis time of approximately 1 hour.
[发布日期] 2011-12-30 [发布机构] 
[效力级别]  [学科分类] 
[关键词] Cardiovascular Magnetic Resonance;Mouse Heart;Cardiac Phase;Myocardial Mechanic;Partition Direction [时效性] 
   浏览次数:1      统一登录查看全文      激活码登录查看全文