Does tibial design modification improve implant stability for total knee arthroplasty? An experimental cadaver study
[摘要] AimsOne of the main causes of tibial revision surgery for total knee arthroplasty is aseptic loosening. Therefore, stable fixation between the tibial component and the cement, and between the tibial component and the bone, is essential. A factor that could influence the implant stability is the implant design, with its different variations. In an existing implant system, the tibial component was modified by adding cement pockets. The aim of this experimental in vitro study was to investigate whether additional cement pockets on the underside of the tibial component could improve implant stability. The relative motion between implant and bone, the maximum pull-out force, the tibial cement mantle, and a possible path from the bone marrow to the metal-cement interface were determined.MethodsA tibial component with (group S: Attune S+) and without (group A: Attune) additional cement pockets was implanted in 15 fresh-frozen human leg pairs. The relative motion was determined under dynamic loading (extension-flexion 20° to 50°, load-level 1,200 to 2,100 N) with subsequent determination of the maximum pull-out force. In addition, the cement mantle was analyzed radiologically for possible defects, the tibia base cement adhesion, and preoperative bone mineral density (BMD).ResultsThe BMD showed no statistically significant difference between both groups. Group A showed for all load levels significantly higher maximum relative motion compared to group S for 20° and 50° flexion. Group S improved the maximum failure load significantly compared to group A without additional cement pockets. Group S showed a significantly increased cement adhesion compared to group A. The cement penetration and cement mantle defect analysis showed no significant differences between both groups.ConclusionFrom a biomechanical point of view, the additional cement pockets of the component have improved the fixation performance of the implant.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 骨科学
[关键词] Attune;Attune S+;Cement debonding;Implant stability;Cement pockets;total knee arthroplasty;tibial components;flexion;metal;tibia;bone mineral density (BMD);Aseptic loosening;Revision surgeries;bone marrow;stable fixation [时效性] 
   浏览次数:4      统一登录查看全文      激活码登录查看全文