[摘要] Late complications can arise after explantation of a left ventricular assist device. We report the case of a patient who presented at the age of 19 years with peri-partum cardiomyopathy and was initially managed with a biventricular support device, which was subsequently upgraded to an ambulatory left ventricular assist device. This was successfully explanted after myocardial recovery via a minimally invasive approach 7 months later. The patient re-presented 5 years following explantation with hemoptysis. At redo sternotomy, a 10-cm remnant of the outflow graft was found to be eroding the surface of the right lung. The conduit was excised and the stump oversewn. Eleven months later, she presented again with hemoptysis, and a pseudo-aneurysm was identified on the computed tomography scan. She underwent urgent open repair with peripheral cannulation and deep hypothermic total circulatory arrest. The Dacron stump was found to be partially dehisced, leaving a 2–3 cm defect in the ascending aorta. The defect was repaired with a bovine pericardial patch. The patient made a good recovery and was successfully discharged.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 
[关键词] Ventricular assist device explantation;Outflow graft pseudo-aneurysm;Hemoptysis;Redo surgery [时效性] 