Performance of human body communication-based wearable ECG with capacitive coupling electrodes
[摘要] Wearable electrocardiogram (ECG) is attracting much attention in daily healthcare applications, and human body communication (HBC) technology provides an evident advantage in making the sensing electrodes of ECG also working for transmission through the human body. In view of actual usage in daily life, however, non-contact electrodes to the human body are desirable. In this Letter, the authors discussed the ECG circuit structure in the HBC-based wearable ECG for removing the common mode noise when employing non-contact capacitive coupling electrodes. Through the comparison of experimental results, they have shown that the authors’ proposed circuit structure with the third electrode directly connected to signal ground can provide an effect on common mode noise reduction similar to the usual drive-right-leg circuit, and a sufficiently good acquisition performance of ECG signals.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 肠胃与肝脏病学
[关键词] electrocardiography;capacitive sensors;circuit noise;biomedical electrodes;body sensor networks;human body communication-based wearable ECG;capacitive coupling electrodes;wearable electrocardiogram;HBC technology;daily life;noncontact electrodes;ECG circuit structure;common mode noise;signal ground;acquisition performance [时效性] 