已收录 268920 条政策
 政策提纲
  • 暂无提纲
Lightweight wrist photoplethysmography for heavy exercise: motion robust heart rate monitoring algorithm
[摘要] The challenge of heart rate monitoring based on wrist photoplethysmography (PPG) during heavy exercise is addressed. PPG is susceptible to motion artefacts, which have to be mitigated for accurate heart rate estimation. Motion artefacts are particularly apparent for wrist devices, for example, a smart watch, because of the high mobility of the arms. Proposed is a low complexity highly accurate heart rate estimation method for continuous heart rate monitoring using wrist PPG. The proposed method achieved 2.57% mean absolute error in a test data set where subjects ran for a maximum speed of 17 km/h.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 肠胃与肝脏病学
[关键词] photoplethysmography;body sensor networks;patient monitoring;watches;medical signal processing;lightweight wrist photoplethysmography;heavy exercise;motion robust heart rate monitoring algorithm;motion artefacts;accurate heart rate estimation;smart watch;low complexity-highly accurate heart rate estimation;mean absolute error;test data set [时效性] 
   浏览次数:3      统一登录查看全文      激活码登录查看全文