Cardiorespiratory system monitoring using a developed acoustic sensor
[摘要] This Letter proposes a wireless acoustic sensor for monitoring heartbeat and respiration rate based on phonocardiogram (PCG). The developed sensor comprises a processor, a transceiver which operates at industrial, scientific and medical band and the frequency of 2.54 GHz as well as two capacitor microphones which one for recording the heartbeat and another one for respiration rate. To evaluate the precision of the presented sensor in estimating heartbeat and respiration rate, the sensor is tested on the different volunteers and the obtained results are compared with a gold standard as a reference. The results reveal that root-mean-square error are determined <2.27 beats/min and 0.92 breaths/min for the heartbeat and respiration rate in turn. While the standard deviation of the error is obtained <1.26 and 0.63 for heartbeat and respiration rate, respectively. Also, the sensor estimate sounds of to obtained PCG signal with sensitivity and specificity 98.1% and 98.3% in turn that make 3% improvement than previous works. The results prove that the sensor can be appropriate candidate for recognising abnormal condition in the cardiorespiratory system.
[发布日期] [发布机构]
[效力级别] [学科分类] 肠胃与肝脏病学
[关键词] acoustic transducers;bioelectric phenomena;pneumodynamics;phonocardiography;cardiorespiratory system monitoring;developed acoustic sensor;wireless acoustic sensor;heartbeat monitoring;respiration rate monitoring;phonocardiogram;PCG;capacitor microphones;frequency 2.54 GHz [时效性]