Ion Selective Electrode (ISE) Method for Determination of Total Fluorine and Total Organic Fluorine in Packaging Substrates
[摘要] 9000 PFAS and are often globally used for regulatory limits. However, this destructive approach requires careful sample preparation, combustion, and the analysis of the solution by a fluoride-specific electrode. The method described herein is a cost-effective, rapid, quantitative, and externally validated initial screening of packaging materials for fluoro-chemistry. This study presents validated protocols for measuring TF and TOF in packaging substrates using oxygen combustion sample preparation coupled with fluoride ion-selective electrode (F-ISE); the materials and required equipment are provided, and the step-by-step procedure from sample preparation to the analysis are described, including critical steps to minimize contamination and interferences during sample preparation.
[发布日期] [发布机构]
[效力级别] [学科分类] 电力
[关键词] per- and polyfluoroalkyl substances (PFAS);food packaging materials;total fluorine;total organic fluorine;ion selective electrode (ISE) [时效性]