Clinical feasibility of remote intermittently scanned continuous glucose monitoring in coronavirus disease 2019 patients with and without diabetes during dexamethasone therapy
[摘要] The clinical utility of intermittently scanned continuous glucose monitoring (isCGM) in patients with coronavirus disease 2019 (COVID-19) is unclear. Hence, we investigated the accuracy of isCGM in COVID-19 patients during dexamethasone therapy. We evaluated the accuracy of the FreeStyle Libre via180 mg/dL, respectively. When divided according to the presence and absence of diabetes, both groups of paired glucose measurements showed a good correlation (56.3% and 43.7%, and 27.1% and 72.9% within the A and B areas in patients with and without diabetes, respectively), but the MARD was not significant but higher in patients without diabetes (16.5% and 24.2% in patients with and without diabetes). In conclusion, although isCGM may not be as accurate as traditional blood glucose monitoring, it has good reliability in COVID-19 patients with and without diabetes during dexamethasone therapy.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 内分泌与代谢学
[关键词] Coronavirus disease 2019;Device accuracy;Intermittently scanned continuous glucose monitoring;Remote continuous glucose monitoring;Dexamethasone [时效性] 