已收录 273081 条政策
 政策提纲
  • 暂无提纲
Maintenance haemodialysis with low dialysate flow rates in Senegal
[摘要] Introduction: The objective of the study reported here was to demonstrate that maintenance haemodialysis using a reduced dialysate flow rate of 300 mL/min (RQ D ) is not inferior to haemodialysis using the standard flow rate of 500 mL/min (SQ D ) in respect of the delivered dose of dialysis.Methods: A prospective, single-centre, sequential study was performed at the haemodialysis  centre of Pikine Hospital in Dakar. Twenty patients were included. During the first week, three haemodialysis sessions were performed with SQ D and during the second week three haemodialysis  sessions were conducted with RQ D for each patient.Results: For SQ D , the mean eKt/V was 1.38 ± 0.58. There were 38 (63%) sessions with eKt/V greater than 1.2 and 16 patients (80%) had adequate dialysis, based on the average eKt/V. For RQ D , the mean eKt/V was 1.2 ± 0.43 with 25 sessions (42%) having an eKt/V greater than 1.2. There were 11 patients (55%) with adequate dialysis. The dialysis dose was higher with the SQ D prescription (P < 0.001). Ten patients with dry weight ≤60 kg had adequate dialysis with RQ D . Cases of hypokalaemia were significantly higher with the SQ D (P = 0.001).Conclusions: RQ D appears to be inferior in terms of dialysis dose. However, for patients with dry weight ≤60 kg, adequate dialysis could be delivered with RQ D, consequently allowing substantial saving of water in haemodialysis.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 数学(综合)
[关键词] low flow dialysate rates;dialysis dose;dry weight;Senegal [时效性] 
   浏览次数:3      统一登录查看全文      激活码登录查看全文