已收录 273628 条政策
 政策提纲
  • 暂无提纲
Comparing the accuracy of point-of-care with laboratory (capillary, venous, and arterial) blood glucose levels in critically ill patients with and without shock
[摘要] ObjectivesTo compare the accuracy of point-of-care capillary and venous/arterial samples to laboratory testing of venous/arterial samples in critically sick shocked and non-shocked patients. This is a prospective case–control study including capillary, venous, and arterial blood samples from 268 critically ill patients. The King Fahd Military Medical Complex in Dhahran, Saudi Arabia, was the site of this investigation.ResultsWe were able to obtain data on 268 patients for this investigation. POCT and lab findings of venous and central blood did not differ significantly (P = 0.389 and 0.208), while POCT indicated somewhat higher results with venous glucose concentrations of 10.18 and 10.05 (POCT and lab tests respectively) and 9.18 and 9.54 (POCT and lab tests respectively). In addition, the mean differences between POC and laboratory analyses of venous, arterial, and central glucose were 0.13, − 1.75, and − 0.36 mmol/L for venous, arterial, and central glucose, respectively. Except for arterial blood glucose, we did not observe a significant difference between POCT and routine laboratory analysis of glucose concentrations in critically ill patients. Compared to laboratory blood analysis, the use of POCT is marginally accurate, with no difference between shocked and non-shocked patients.
[发布日期] 2022-11-24 [发布机构] 
[效力级别]  [学科分类] 
[关键词] Blood glucose;Intensive care unit;Patients in shock;Laboratory [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文