已收录 273594 条政策
 政策提纲
  • 暂无提纲
Results Of The Excreta Bioassay Quality Control Program For April 1, 2007 Through March 31, 2008
[摘要] A total of 79 urine samples, 3 blank fecal and 5 spiked artificial fecal samples were submitted during the report period (April 1, 2007 through March 31, 2008) to General Engineering Laboratories, South Carolina by the Hanford Internal Dosimetry Program (IDP) to check the accuracy, precision, and detection levels of their analyses. Urine analyses for tritium, Sr, 238Pu, 239Pu, 241Am, 243Am 235U, 238U, elemental uranium and fecal analyses for 241Am, 238Pu and 239Pu were tested this year. The number of QC urine samples submitted during the report period represented 1.8% of the total samples submitted. In addition to the samples provided by IDP, GEL was also required to conduct their own QC program, and submit the results of analyses to IDP. About 35% of the analyses processed by GEL during the third year of this contract were quality control samples. GEL tested the performance of 24 radioisotopes, all of which met or exceeded the specifications in the Statement of Work within statistical uncertainty (Table 4). IDP concluded that GEL was performing well for all analyses tested, and concerns identified earlier were satisfactorily resolved.
[发布日期] 2008-12-31 [发布机构] 
[效力级别]  [学科分类] 核能源与工程
[关键词] FECES;BIOASSAY;DOSIMETRY;PERFORMANCE;QUALITY CONTROL;TRITIUM;URANIUM ISOTOPES;URINE;PLUTONIUM ISOTOPES;AMERICIUM ISOTOPES;STRONTIUM excreta bioassay;urine analyses;fecal analyses [时效性] 
   浏览次数:45      统一登录查看全文      激活码登录查看全文