The Gas Flow from the Gas Attenuator to the Beam Line
[摘要] The gas leak from the gas attenuator to the main beam line of the Linac Coherent Light Source has been evaluated, with the effect of the Knudsen molecular beam included. It has been found that the gas leak from the gas attenuator of the present design, with nitrogen as a working gas, does not exceed 10{sup -5} torr x l/s even at the highest pressure in the main attenuation cell (20 torr).
[发布日期] 2010-12-03 [发布机构]
[效力级别] [学科分类] 核物理和高能物理
[关键词] ATTENUATION;GAS FLOW;LIGHT SOURCES;LINEAR ACCELERATORS;LEAKS;NITROGEN;FREE ELECTRON LASERS XFEL [时效性]