Measurement of six degree motion error of linear stage
[摘要] In this paper, we propose measurement system of the six degree of motion errors which is based on distance measurement by the laser interferometer. The system has six parallel laser beams and six corner cube mirrors on the linear stage, which reflect the corresponding laser beams. The error of axial direction is measured with the ordinary distance measurement method by laser interferometer. The vertical errors to the axial direction and the roll errors are measured by tilted beams using the wedge prism. The pitch and yaw errors are measured by the difference between distance of two corner cube mirrors. The experimental layout of the corner cube mirrors and the other optical devices are shown. As a result, the resolution of 66.5 nm for error of the axial direction, the resolution of 383 nm for the vertical errors to the axial direction, the resolution of 1.23 arc-sec for the roll error and the resolution of 0.429 arc-sec for the pitch and yaw errors are obtained in this system.
[发布日期] [发布机构] Tokyo Denki University, School of Engineering, Department of Mechanical Engineering, Senju-Asahi 5, Adachi, Tokyo; 120-8551, Japan^1
[效力级别] 机械制造 [学科分类] 航空航天科学
[关键词] Axial direction;Corner cubes;Laser interferometer;Linear stages;Measurement system;Motion errors;Tilted beams;Wedge prisms [时效性]