Raman Under Liquid Nitrogen (RUN)
[摘要] The recording of Raman spectra for many molecules in air at room temperature is difficult or impossible as a result of sample degradation which is due to a combination of laser heating and oxidation. Often nitrogen gas is applied over the sample in an attempt to reduce oxidation. Also, the samples are sometimes cooled to reduce ro-vibrational "hot bands" and enhance the spectra. We have found great utility in recording Raman spectroscopy of samples under liquid nitrogen, a technique we call RUN. The RUN spectra show much higher resolution as a result of ro-vibrational cooling and in some cases cooling produces only the lowest energy conformer of the molecular ensemble further simplifying the spectra. A very sharp Raman peak at 2327.0 cm-1, due to liquid nitrogen, also serves as a convenient wavelength calibration. We also demonstrate the ability to clearly delineate the lattice modes for naphthalene and benzene crystals.
[发布日期]  [发布机构] Departments of Physics and Chemistry, University of Tennessee, Knoxville; TN; 37996, United States^1;Department of Chemistry and Biochemistry, University of Mississippi, PO Box 1848, University; MS; 38677, United States^2
[效力级别]  [学科分类] 
[关键词] Combination of lasers;Energy conformers;Higher resolution;Lattice modes;Nitrogen gas;Raman peak;Vibrational cooling;Wavelength calibration [时效性] 