A small animal image guided irradiation system study using 3D dosimeters
[摘要] In a high resolution image-guided small animal irradiation platform, a cone beam computed tomography (CBCT) is integrated with an irradiation unit for precise targeting. Precise quality assurance is essential for both imaging and irradiation components. The conventional commissioning techniques with films face major challenges due to alignment uncertainty and labour intensive film preparation and scanning. In addition, due to the novel design of this platform the mouse stage rotation for CBCT imaging is perpendicular to the gantry rotation for irradiation. Because these two rotations are associated with different mechanical systems, discrepancy between rotation isocenters exists. In order to deliver x-ray precisely, it is essential to verify coincidence of the imaging and the irradiation isocenters. A 3D PRESAGE dosimeter can provide an excellent tool for checking dosimetry and verifying coincidence of irradiation and imaging coordinates in one system. Dosimetric measurements were performed to obtain beam profiles and percent depth dose (PDD). Isocentricity and coincidence of the mouse stage and gantry rotations were evaluated with starshots acquired using PRESAGE dosimeters. A single PRESAGE dosimeter can provide 3 -D information in both geometric and dosimetric uncertainty, which is crucial for translational studies.
[发布日期] [发布机构] Department of Radiation Oncology, Columbia University, New York; NY, United States^1;Department of Chemistry and Biology, Rider University, Lawrencevillea; NJ, United States^2;Heuris Pharma, LLC, 412 Sunset Rd., Skillman; NJ; 08558, United States^3
[效力级别] [学科分类]
[关键词] 3d dosimeters;3D information;Beam profiles;Cone-beam computed tomography;High resolution image;Irradiation system;Labour-intensive;Mechanical systems [时效性]