已收录 268921 条政策
 政策提纲
  • 暂无提纲
Manufacturability of lab on chip devices : reagent-filled reservoirs bonding process and its effect on reagents flow pattern
[摘要] In its lab-on-a-chip product, Daktari Diagnostics utilizes ;;reagent-filled reservoirs;; as a means of storing and delivering the liquid reagent. During the clinical trials of the product, undesired reagent flow patterns (namely, flow anomaly 1 & flow anomaly 2) were noticed. This work focused on optimizing the bonding process of the reagent-filled reservoirs to the backbone. Also, the relationship between the bonding process parameters and the reagent flow pattern was studied in depth. To achieve the objective of this thesis, an experiment was designed in which independent variables were the heat sealing parameters (x, y, z) and accelerated aging and the dependent variables were bond strength and the reagent flow pattern. Experiments showed that optimal Heat Sealing parameters are: parameter x = 4.5, parameter y = 110 and parameter z = 1.5. At the optimal settings of bonding process, the highest bond strength was attained and the reagent flow improved considerably but flow anomalies were not completely resolved. Also, results showed that accelerated aging affected the bond strength negatively. Accelerated aging also affected the flow pattern negatively, but this effect was not statistically significant.
[发布日期]  [发布机构] Massachusetts Institute of Technology
[效力级别]  [学科分类] 
[关键词]  [时效性] 
   浏览次数:3      统一登录查看全文      激活码登录查看全文