已收录 272903 条政策
 政策提纲
  • 暂无提纲
Temperature-Corrected Oxygen Detection Based on Multi-Mode Diode Laser Correlation Spectroscopy
[摘要] Temperature-corrected oxygen measurements were performed by using multi-mode diode laser correlation spectroscopy at temperatures ranging between 300 and 473 K. The experiments simulatein situmonitoring of oxygen in coal-combustion exhaust gases at the tail of the flue. A linear relationship with a correlation coefficient of−0.999was found between the evaluated concentration and the gas temperature. Temperature effects were either auto-corrected by keeping the reference gas at the same conditions as the sample gas, or rectified by using a predetermined effective temperature-correction coefficient calibrated for a range of absorption wavelengths. Relative standard deviations of the temperature-corrected oxygen concentrations obtained by different schemes and at various temperatures were estimated, yielding a measurement precision of 0.6%.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 光谱学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文