Diode-Laser Induced Fluorescence Spectroscopy of an Optically Thick Plasma in Combination with Laser Absorption Spectroscopy
[摘要] Distortion of laser-induced fluorescence profiles attributable to optical absorption and saturation broadening was corrected in combination with laser absorption spectroscopy in argon plasma flow. At high probe-laser intensity, saturated absorption profiles were measured to correct probe-laser absorption. At low laser intensity, nonsaturated absorption profiles were measured to correct fluorescence reabsorption. Saturation broadening at the measurement point was corrected using a ratio of saturated to non-saturated broadening. Observed LIF broadening and corresponding translational temperature without correction were, respectively,2.20±0.05 GHz and2510±100 K and corrected broadening and temperature were, respectively,1.96±0.07 GHz and1990±150 K. Although this correction is applicable only at the center of symmetry, the deduced temperatureagreed well with that obtained by LAS with Abel inversion.
[发布日期] [发布机构]
[效力级别] [学科分类] 光谱学
[关键词] [时效性]