已收录 273176 条政策
 政策提纲
  • 暂无提纲
Validated Reverse-Phase High-Performance Liquid Chromatography for Quantification of Furosemide in Tablets and Nanoparticles
[摘要] A simple, sensitive, and specific method for furosemide (FUR) analysis by reverse-phase-HPLC was developed using a Spherisorb C18ODS 2 column. A chromatographic analysis was carried out using a mobile phase consisting of acetonitrile and 10 mM potassium phosphate buffer solution: 70 : 30 (v/v) at pH 3.85, at a flow rate of 1 mL·min−1. The UV-detection method was carried out at 233 nm at room temperature. Validation parameters including limit of detection (LOD), limit of quantitation (LOQ), linearity range, precision, accuracy, robustness, and specificity were investigated. Results indicated that the calibration curve was linear (r2=0.9997) in the range of 5.2 to 25,000 ng·mL−1, withεvalue equal to3.74×104 L·M−1·cm−1. The LOD and LOQ were found to be 5.2 and 15.8 ng·mL−1, respectively. The developed method was found to be accurate (RSD less than 2%), precise, and specific with an intraday and interday RSD range of 1.233–1.509 and 1.615 to 1.963%. The stability of native FUR has also been performed in simulated perilymph and endolymph media (with respective potency in each medium of99.8±2.3% and96.68±0.7%,n=3) after 6 hours. This method may be routinely used for the quantitative analysis of FUR from nanocarriers, USP tablets and release media related to hearing research
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文