A Stability-Indicating HPLC-DAD Method for Determination of Ferulic Acid into Microparticles: Development, Validation, Forced Degradation, and Encapsulation Efficiency
[摘要] A simple stability-indicating HPLC-DAD method was validated for the determination of ferulic acid (FA) in polymeric microparticles. Chromatographic conditions consisted of a RP C18column (250 mm × 4.60 mm, 5 μm, 110 Å) using a mixture of methanol and water pH 3.0 (48 : 52 v/v) as mobile phase at a flow rate of 1.0 mL/min with UV detection at 320 nm. The developed method was validated as per ICH guidelines with respect to specificity, linearity, limit of quantification, limit of detection, accuracy, precision, and robustness provided suitable results regarding all parameters investigated. The calibration curve was linear in the concentration range of 10.0–70.0 μg/mL with a correlation coefficient >0.999. Precision (intraday and interday) was demonstrated by a relative standard deviation lower than 2.0%. Accuracy was assessed by the recovery test of FA from polymeric microparticles (99.02% to 100.73%). Specificity showed no interference from the components of polymeric microparticles or from the degradation products derived from acidic, basic, and photolytic conditions. In conclusion, the method is suitable to be applied to assay FA as bulk drug and into polymeric microparticles and can be used for studying its stability and degradation kinetics.
[发布日期] [发布机构]
[效力级别] [学科分类] 分析化学
[关键词] [时效性]