1H MRS phantom studies of BNCT10B‒carrier, BPA–F using STEAM and PRESS MRS sequences: Detection limit and quantification
[摘要] The quantification of boron neutron capture therapy (BNCT)10B‒carrier, L‒p‒boronophenylalanine‒fructose complex (BPA–F) was studied with phantoms using1H magnetic resonance spectroscopy sequences PRESS and STEAM at 1.5 and 3.0 T. The results show that typical attainable short echo times of clinical MRS sequences combined with long repetition time result in clinically acceptable quantification accuracy. However, the concentration ratios, which are essential for the treatment planning, can still be reliably measured by using small repetition times. Detection limits of BPA in aqueous phantoms at 1.5 and 3.0 T were evaluated using clinically acceptable measurement time of ~10 min, two typical voxel sizes (153and 203mm3) and PRESS and STEAM sequences. The detection limits of BPA in phantom conditions were 0.7 (3.0 T) and 1.4 mM (1.5 T) for PRESS sequence with 203mm3voxel.