Internal Mammary Arterial Injury from Lead Extraction: A Clinically Subtle yet Important Complication of Implantable Device Removal
[摘要] Percutaneous implantable device extraction has increased in recent years and is associated with small but significant risk. Arteriovenous fistula formation is an uncommon complication of this procedure.We report two cases where lead extraction was complicated by an arteriovenous fistula between the left internal mammary artery and the left brachiocephalic vein. In both cases, the patients were asymptomatic and the presence of a continuous murmur in the left subclavicular region led to the appropriate diagnosis. These were successfully treated with coil embolization.Auscultation around prior extraction sites should be routinely done to aid in the diagnosis of this potentially harmful complication.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 心脏病和心血管学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文