[摘要] A commercially available fibrinogen standard calibrated by a World Health Organization (WHO) reference material is widely used in Japan, and most clinical laboratories use the Clauss method for plasma fibrinogen measurement. However, a current issue in fibrinogen measurement is poor laboratory-to-laboratory variability. To improve the reliability of fibrinogen values and thereby solve the poor precision and accuracy of plasma fibrinogen testing, the present paper develops a simple and large preparation procedure for a suitable fibrinogen standard and quality control material and evaluates their basic performance. With a new procedure getting high purified fibrinogen by glycine precipitation, the calibrator determined by both the Clauss and Jacobson methods produced a fibrinogen concentration of 2.20 g l−1. The total precision of the calibrator was excellent (coefficient of variation 1.4-2.1%) in comparison with current plasma fibrinogen materials from the WHO (#98/612) and with a commercial standard (CV 1.9-3.9%). The within-run precision of the calibrator on the coagulation analysers was 1.7-2.8%. Within-analyser variability among the five instruments had good consistency (mean 2.20 ± 0.022 g l−1; CV 1.0%). The degradation study of the calibrator suggested that storage at 9°C for two years was as predicted. In conclusion, the results show that the calibrator prepared herein can be useful as a candidate Japanese fibrinogen standard and is applicable to automated and semi-automated coagulation analysers. Additionally, it is expected that it will be widely used in Japan by diagnostic manufacturers and clinical laboratories as a recommended secondary standard to estimate a fibrinogen value according to the WHO primary standard.
[发布日期] [发布机构]
[效力级别] [学科分类] 分析化学
[关键词] [时效性]