An Environmental Friendly Procedure for Photometric Determination of Hypochlorite in Tap Water Employing a Miniaturized Multicommuted Flow Analysis Setup
[摘要] A photometric procedure for the determination ofClO-in tap water employing a miniaturized multicommuted flow analysis setup and an LED-based photometer is described. The analytical procedure was implemented using leucocrystal violet (LCV;4,4′,4′′-methylidynetris (N,N-dimethylaniline), C25H31N3) as a chromogenic reagent. Solenoid micropumps employed for solutions propelling were assembled together with the photometer in order to compose a compact unit of small dimensions. After control variables optimization, the system was applied for the determination ofClO-in samples of tap water, and aiming accuracy assessment samples were also analyzed using an independent method. Applying the pairedt-test between results obtained using both methods, no significant difference at the 95% confidence level was observed. Other useful features includelow reagent consumption, 2.4 μg of LCV per determination, a linear response ranging from 0.02 up to 2.0 mg L−1 ClO-, a relative standard deviation of 1.0% (n=11) for samples containing 0.2 mg L−1 ClO-, a detection limit of 6.0 μg L−1 ClO-, a sampling throughput of 84 determinations per hour, and a waste generation of 432 μL per determination.
[发布日期] [发布机构]
[效力级别] [学科分类] 分析化学
[关键词] [时效性]