已收录 272983 条政策
 政策提纲
  • 暂无提纲
Design of LabVIEW®-based software for the control of sequential injection analysis instrumentation for the determination of morphine
[摘要] LabVIEW®-based software for the automation of a sequential injection analysis instrument for the determination of morphine is presented. Detection was based on its chemiluminescence reaction with acidic potassium permanganate in the presence of sodium polyphosphate. The calibration function approximated linearity (range5×10−10to5×10−6M) with a line of best fit ofy=1.05x+89164(R2=0.9959), where y is the log10 signal (mV) and x is the log10 morphine concentration (M). Precision, as measured by relative standard deviation, was 0.7% for five replicate analyses of morphine standard (5×10−8M). The limit of detection (3σ) was determined as5×10−11M morphine.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:4      统一登录查看全文      激活码登录查看全文