Survival Analysis of Factors Influencing Cyclic Fatigue of Nickel-Titanium Endodontic Instruments
[摘要] Objective. The aim of this study was to validate a survival analysis assessing the effect of type of rotary system, canal curvature, and instrument size on cyclic resistance.Materials and Methods. Cyclic fatigue testing was carried out in stainless steel artificial canals with radii of curvature of 3 or 5 mm and the angle of curvature of 60 degrees. All the instruments were new and 25 mm in working length, and ISO colour coding indicated the instrument size (yellow for size 20; red for size 25).Wizard Navigatorinstruments,Mtwoinstruments,ProTaperinstruments, andRevo-Sinstruments were passively rotated at 250 rotations per minute, and the time fracture was being recorded. Subsequently, fractographic analysis of broken tips was performed by scanning electron microscope. The data were then analysed by the Kaplan-Meier estimator of the survival function, the Cox proportional hazards model, the Wald test for regression covariates, and the Wald test for significance of regression model.Conclusion. The lifespan registered for the tested instruments wasMtwo>Wizard Navigator>Revo-S>ProTaper; 5 mm radius > 3 mm radius; and yellow > red in ISO colour coding system.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 材料工程
[关键词]  [时效性] 