Assessing the ABL 500 blood gas analyser
[摘要] The ABL 500 blood gas analyser from Radiometer has cordlesselectrodes and does not use a humidifier for calibrating gases. During the evaluation of the analytical performance of this instrument, the problem of p02accuracy was approached by comparing the values obtained with two kinds of tonometry (film and bubble). An acceptable level of imprecision was demonstrated for all measured parameters. For within-run precision, with tonometry, coefficients of variation (CV) were ≆0.37% for pO2and ≆0.52% for pCO2. A CV of 1.76% was found for day-to-day precision for both p02and pCO2. In the linearity study, with both tonometry methods, and in the inter-instrument comparisons (the ABL was compared with the Ciba Corning 178), pO2valuesobtained on the ABL 500 exhibited a slight overestimation above 150 mmHg (2.2-3.4% at 600 mmHg). This minor discrepancy is discussed with reference to the new design of the pO2electrode, the algorithm for pO2correction and the tonometry procedure. The results reported in this paper stress the importance of pO2accuracy assessment for the evaluation of blood gas analysers.