已收录 272378 条政策
 政策提纲
  • 暂无提纲
Automated cold vapour flow-injection analysis of mercury at high concentrations
[摘要] Continuous-flow cold vapour- atomic fluorescence spectrometry isshown to be an extremely sensitive technique for the determination of mercury with detection limits typically below 0.01 μg l-1. Linear calibration ranges were found to be at least four orders of magnitude (i.e. up to 0.1 mg l-1). Samples with concentrations exceeding the linear range are susceptible to self-absorption, and may, in severe cases, cause carry-over problems between samples. The flow-injection approach has been utilized to extend the upper limit of the linear calibration range allowing determinations up to 10 mg l-1of mercury. A range of certified reference materials and zinc battery anodes have been successfully analysed with a minimalnumber of sample dilutions.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文