已收录 272983 条政策
 政策提纲
  • 暂无提纲
Performance of a reflectometric techniquefor serum alanine aminotransferasedeterminations
[摘要] The Ames Seralyzer uses solid phase chemistry coupled withreflectance spectroscopy to measure levels of serum alanineaminotransferase (serum ALT). The study reported here wascarried out to evaluate the performance of the Seralyzer methodcompared to conventional analyses carried out on a Coulter Dacosdiscrete random access analyser using Boehringer (BCL) reagentsconforming to SCE specications at 37° C. Clinically acceptableaccuracy and precision were obtained. The effect of bilirubin andhaemoglobin on the test was investigated. Increased levels ofbilirubin in sera did not affect the estimation but the presence ofhigh haemoglobin levels produced significant interference. Theoverall practicality of the Seralyzer was assessed and the methodwas found to be particularly suitable for the estimation of urgent orhigh risk samples in the laboratory or for use by suitably trainedpersonnel in other clinical areas.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 分析化学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文