Evaluation of the IL-Phoenix chemistry electrolyte analyser
[摘要] This paper reports an evaluation of the IL-Phoenix Chemistry/Electrolyte Analyser; the evaluation was carried out in accordance with internationally recognized guidelines. The evaluation was performed in three steps: evaluation in routine conditions; assessment of interferences; and study of practicability. Seven constituents were studied under routine working conditions. Within-run imprecision rangedfrom 0.6% (CV) for chloride to 3.1% (CV) for glucose. Between-run imprecision ranged from 0.9% for sodium to 6.0% (CV) for urea. Sample-related carryover was not significant. The relative inaccuracy was acceptable; drift was negligible; linearity was agreed with the range showed by the supplier. Haemoglobin produced negative interferences with sodium and chloride. Turbidity interfered negatively with sodium, chloride, potassium and total calcium, andpositively with glucose. Bilirubin showed a negative interference with sodium, chloride and creatinine.