已收录 268921 条政策
 政策提纲
  • 暂无提纲
Biomechanical analysis of acromioclavicular joint dislocation repair using coracoclavicular suspension devices in two different configurations
[摘要]

Background

The best treatment option for some acromioclavicular (AC) joint dislocations is controversial. For this reason, the aim of this study was to evaluate the vertical biomechanical behavior of two techniques for the anatomic repair of coracoclavicular (CC) ligaments after an AC injury.

Materials and methods

Eighteen human cadaveric shoulders in which repair using a coracoclavicular suspension device was initiated after injury to the acromioclavicular joint were included in the study. Three groups were formed; group I (n = 6): control; group II (n = 6): repair with a double tunnel in the clavicle and in the coracoid (with two CC suspension devices); group III (n = 6): repair in a “V” configuration with two tunnels in the clavicle and one in the coracoid (with one CC suspension device). The biomechanical study was performed with a universal testing machine (Electro Puls 3000, Instron, Boulder, MA, USA), with the clamping jaws set in a vertical position. The force required for acromioclavicular reconstruction system failure was analyzed for each cadaveric piece.

Results

Group I reached a maximum force to failure of 635.59 N (mean 444.0 N). The corresponding force was 939.37 N (mean 495.6 N) for group II and 533.11 N (mean 343.9 N) for group III. A comparison of the three groups did not find any significant difference despite the loss of resistance presented by group III.

Conclusion

Anatomic repair of coracoclavicular ligaments with a double system (double tunnel in the clavicle and in the coracoid) permits vertical translation that is more like that of the acromioclavicular joint. Acromioclavicular repair in a “V” configuration does not seem to be biomechanically sufficient.

[发布日期]  [发布机构] 
[效力级别]  [学科分类] 骨科学
[关键词] Biomechanics [时效性] 