Automated Reaction-Rate Method for Determination of Serum Creatinine with the CentrifiChem
[摘要] An automated kinetic method to determine creatinine in serum with the Jaffé reaction is described. The method does not require deproteinization or adsorption of creatinine on Fuller's earth or ion-exchange resin. A maximum of 29 serum samples, standards, and control sera can be assayed in less than 1.5 min. Interferences by most of the known Jaffé-positive serum constituents were studied and minimized by optimal selection of the time interval during which the reaction rate was measured. Results of precision studies and correlation with a manual method requiring adsorption of creatinine on ion-exchange resin are reported. Spectroscopic data suggest that the product of reaction between picrate and creatinine is a 1:1 adduct of the two starting molecules.
[发布日期] [发布机构]
[效力级别] [学科分类] 过敏症与临床免疫学
[关键词] [时效性]