Rapid, Reliable Method for Measuring Serum Lipase Activity
[摘要] The method of Dirstine et al. [Clin. Chem. 14, 1097 (1968)] has been modified to increase sensitivity and decrease manipulations. The modifications permit a 10-min incubation, and a single analysis can be completed in 30 min, or five concomitant assays can be done in an hour. Stability of the substrate has been increased by increasing the emulsifier concentration. Recovery of the fatty acid products has been improved by adding electrolyte to the copper reagent and by using combined organic solvents. Pancreatin was added to human serum to provide samples with increased activities, with behavior identical to that of pathological sera with increased lipase activities. Normal values for 100 adults, 50 women and 50 men, ranged between 21 and 171 units, with a mean of 63 units. Within-day precision for 21 assays (mean activity, 641 units) was ±16 units (SD). Between-days precision for 36 assays was ±18 units (SD), with a mean of 636 units.
[发布日期] [发布机构]
[效力级别] [学科分类] 过敏症与临床免疫学
[关键词] [时效性]