Europium and samarium as labels in time-resolved immunofluorometric assay of follitropin.
[摘要] This time-resolved immunofluorometric assay for human follitropin involves use of europium- or samarium-labeled monoclonal antibodies, with an average incorporation ratio of 3 mol of Eu3+ or Sm3+ per mole of antibody. These lanthanide ions are bound to the antibody molecules by means of the anhydride of diethylenetriaminepentaacetic acid. The solid-phase antibody is immobilized inside polystyrene tubes in which plasma samples were assayed in a one-step procedure. After incubation, the fluorescence intensity of Eu3+ or Sm3+ label is measured by time-resolved fluorometry, with a nitrogen laser as the pulsed excitation source. The sensitivity of the assay is largely better with Eu3+ than with Sm3+ because of the difference in their intrinsic luminescence properties. Results obtained with the proposed methods correlated well with those by an immunoradiometric method.
[发布日期]  [发布机构] 
[效力级别]  [学科分类] 过敏症与临床免疫学
[关键词]  [时效性] 
   浏览次数:2      统一登录查看全文      激活码登录查看全文