Electrochemical determination of low blood lead concentrations with a disposable carbon microarray electrode.
[摘要] Lead concentrations in blood samples typical of unexposed individuals are determined by using square-wave anodic stripping voltammetry at a disposable 287-element (15-35-microns diameter) carbon microarray electrode. Analysis of a series of low Pb-containing samples (blood [Pb] = 17 to 92 micrograms/L) by standard addition gives good accuracy (average bias vs graphite furnace atomic absorption spectroscopy = -4 micrograms/L) and precision (pooled SD = 7 micrograms/L). The signal/noise ratio is improved by about an order of magnitude over that found at a single carbon disk (10 microns diameter), resulting in a detection limit (for the microarray) of 5 micrograms/L for a 60-s deposition of sample.
[发布日期] [发布机构]
[效力级别] [学科分类] 过敏症与临床免疫学
[关键词] [时效性]