Adaptation of the EMIT serum digoxin assay to a mini-disc centrifugal analyzer.
[摘要] We have modified the EMIT (Syva) serum digoxin assay for use with a mini-disc centrifugal analyzer. A minicomputer is used to control data acquisition, calculations, and readout for 14 cuvettes simultaneously. Thus, it is not necessary for time each individual assay manually or to plot calibration curves on graph paper. We use incubation times and dilutions similar to those used for the manual procedure, but we have changed the buffer solution, decreased the measurment time, and decreased the required serum volume per assay fivefold. We evaluated day-to-day and within-run precision, using Syva calibrators and control sera. EMIT assay values for radioimmunoassay control sera correlate well with the supplier's stated values.
[发布日期] [发布机构]
[效力级别] [学科分类] 过敏症与临床免疫学
[关键词] [时效性]