Optimum Kinetic Enzymatic Procedures for Glucose and Triglycerides in Plasma and Serum
[摘要] I describe optimum kinetic procedures for measuring glucose and triglycerides in plasma and serum by enzymatic techniques. Glucose was measured by the hexokinase method at a 1000-fold sample dilution. Absorbance and concentration are linearly related for concentrations up to 4 g/liter. Results are reproducible to about 25 mg/liter (standard error). Comparisons between the kinetic glucose method described here and an automated o -toluidine and a manual end-point method showed no apparent bias between the methods; correlation coefficients were 0.998 and 0.991, respectively. Triglycerides (triacylglycerols) were measured in a fully enzymatic system by measuring free glycerol after hydrolysis with lipase. Absorbance and concentration are linearly related to >2.5 g/liter at a 300-fold sample dilution. Repeatability was about 30 mg/liter (standard error). Compared with a manual method, the correlation coefficient was 0.978, with a slope of 0.93.